Specifications
Motor power0.1kW, single phase
Maximum input
particle size
10mm
Width x Height x Depth600 x 557 x 496mm
Weight85kg
Hopper capacity5kg
Dust extractionNA
OptionsNA
WarrantyRocklabs Bench Top RSD is sold with a 2000 operating hours or one year warranty
(whichever comes first), covering all parts (excluding wear parts) and labour if used under laboratory conditions.
Comments on taking a representative split
Taking a representative split from a large sample is not as simple as many people believe. Rolling on a mat, multiple grabs or riffle splitting do not produce accurate results. A classic study* conducted at Bradford University demonstrated that a rotating sample divider provided the best results.

*”Critical evaluation of powder sample procedures” . Allen T& Khan A.A “ The chemical Engineer”, May 1979, pp108 - 112
Dividing
MethodStandard deviation of samples %
Cone and quartering6.810
Scoop sampling5.140
Riffle splitting1.010
Table sampling2.090
Rotating splitting0.125
Random variation0.076
Shipping Dimensions
Width825mm
Height730mm
Depth605mm
Weight100kg

*For more information please contact us list of international agents.